(SANEPR.com) June 10, 2009 -- Abstract: Based on standard ASTM D 3985, this article briefs on coulometric oxygen permeability testing of film and sheet materials. It also presents a detailed introduction about the principal and operating procedure of coulometric testing.
Keywords: ASTM D 3985, coulometric sensor, www.labthink.cn, transmission rate, Faraday's law, and coulometric analysis

For more articles, please visit: www.labthink.cn.
With the accelerating development of trade globalization, some testing technologies used in American commerce and trade gradually appeared in our country. After China entering into WTO, two sensor permeability testing methods (coulometric sensor method for film oxygen permeability testing and electrolytic method for film water vapor permeability testing) are added to the testing standards of medicine package. Since these two methods are not methods adopted by national standards, they are less applied in our country.
Introduction of coulometric method is rather rare in YBB00082003. The following is a detailed introduction on the standard ASTM D 3985, which is the reference standard of YBB00082003.
1. Application Fields
This test method covers a procedure for determination of the steady-state rate of transmission of oxygen gas through plastics in the form of film, sheeting, laminates, coextrusions, or plastic-coated papers or fabrics. It only applies to oxygen permeability testing of specimen in dry status (Relative moisture less than1% RH). Test performed under wet condition can refer to ASTM F 1927. Oxygen permeability of most film will vary greatly with the changing of humidity.
2. Summary of Testing Method
The specimen is mounted as a sealed semi-barrier between two chambers at ambient atmospheric pressure. One chamber is slowly purged by a stream of nitrogen and the other chamber contains oxygen. As oxygen gas permeates through the film into the nitrogen carrier gas, it is transported to the coulometric detector where it produces an electrical current, the magnitude of which is proportional to the amount of oxygen flowing into the detector per unit time. Among that, oxygen gas is the testing gas and nitrogen gas is the carrier gas. Oxygen gas concentration of upper chamber is higher than that of lower chamber, due to which certain concentration difference is formed between two sides of specimen. During the permeability process, oxygen gas transmits from upper chamber through specimen into lower chamber.
3. Testing Interference
The presence of certain interfering substances in the carrier gas stream may give rise to unwanted electrical outputs and error factors. Interfering substances include free chlorine and some strong oxidizing agents. Exposure to carbon dioxide should also be minimized to avoid damage to the sensor through reaction with the potassium hydroxide electrolyte.
4. Instrument Calibration
The oxygen sensor used in this test method is a coulometric device that yields a linear output as predicted by Faraday's Law. In principle, four electrons are produced by the sensor for each molecule of oxygen that passes into it. Experience has shown… under some circumstances the sensor may become depleted or damaged to the extent that efficiency and response are impaired. For that reason, this test method incorporates means for a periodic sensor evaluation. standard film is needed in instrument calibration. Since the data of standard film can directly influence the determination of calibration coefficient Q, special attention should be paid to film preparation and calibration repeatability. In this standard, NIST standard material is used for calibration testing.
5. Testing Process
Preparation of apparatus… Heating the apparatus will speed the drying and outgassing process… Inserting the specimen… Purging the system… purge air from the upper and lower diffusion cell chambers, using a flow rate of 50 to 60 mL. After 3 or 4 min, reduce the flow rate to the desired value between 5 and 15 mL/min. Maintain this configuration for 30min… INSERT THE SENSOR so that the carrier gas which has passed through both sides of the diffusion cell is diverted into the sensor… Establish E 0 … Switch OXYGEN into the test-gas side of the diffusion cell… The sensor output current should increase gradually, ultimately stabilizing at a constant value. While some thin films with high diffusion coefficients may reach equilibrium in 30 to 60 min, thicker or more complex structures may require a number of hours to reach a steady state of gas transmission. The steady-state voltage value of the oxygen transmission rate shall be recorded and labelled Ee. Appropriateness of carrier gas flow velocity is of critical importance in the testing, which must be adjusted accurately according to the standard.
Oxygen diffusion coefficient can be calculated according to the formula in standard with the E0 and Ee obtained during testing.
6. Other Cautions
First, oxygen sensor is of consumption type. The sensor should be replaced when there is obvious attenuation of output signal and it cannot be offset by calibration. Back diffusion of air into the unit is undesirable, because oxygen in air will deplete the sensor and therefore shorten the lifespan.
Secondly, High oxygen concentrations in the carrier gas, from the testing of poor barriers, will tend to produce detector saturation. This will shorten the lifespan of sensor. However, this problem can be solved through two ways. One is to reduce oxygen concentration (reduce concentration difference between two sides of the specimen) in testing gas by mixing oxygen and nitrogen. Another way is to reduce specimen surface area through sticking barrier board made of metal or aluminum foil to two sides of specimen with adhesive or ethoxyline resin. In this way, specimen area is equal to the opening area on barrier board.
Thirdly, testing temperature should be controlled. Temperature is a critical parameter affecting the measurement of O2GTR. Strict temperature controlling can reduce the influence of temperature fluctuation on testing results.
7. Conclusions
ASTM D 3985 is the commonly used standard for film oxygen permeability testing using oxygen sensor at present. It is also the reference standard of coulometric analysis method in YBB00082003. A deep understanding of this standard will help the better implementation of YBB00082003. Temperature and humidity are critical parameters influencing final test data. That is why efficiency controlling of testing environment is a way to improve testing precision and efficiency.
###
About Labthink:
Labthink ( www.labthink.cn ), the excellent provider of testing instruments and testing services, is devoted to provide most excellent and complete quality control solutions for pharmaceutical, foodstuff, cosmetics, packaging, printing, adhesive, automotive, petrochemistry, environment, biology, new energy, construction, aviation and electronic industries worldwide.
Labthink Instruments Co., Ltd.
Mark Won Secretary@labthink.cn
Tel: +86 (0)531 85864214-8058
Fax: +86 (0)531 85812140
www.labthink.cn

Add to SimpySimpy Add to Google BookmarksGoogle Add to your Technorati favoritesTechnorati Add to del.icio.usdel.icio.us Add to Redditreddit Add to DiggDigg Add to FurlFurl It Add to SpurlSpurl Add to BlinklistBlink It Add to MyYahoo!MyYahoo Add to Newsvine.comNewsvine

TechnoratiTechnorati Tags:

AddThis Social Bookmark Button

Related Press Releases: 
nCircle Receives Network Products Guide 2010 Product Innovation Award 20/03/2010
SMA to Attend Symitar's Midwest User's Group in East Lansing, MI 19/03/2010
Cine-tal Systems to Include Blue Echo Solutions hd EZ lock with Davio Visual Display Processors 19/03/2010
Mobile Sales Trend Of Today 14/03/2010
Verizon will Issue 4G Cell Phone in the Next Midyear 12/03/2010
SMA attends Symitar's Southwest User's Group Conference 12/03/2010
NovaStor Launches Storageline.com – a SaaS Connection Portal for High­Quality Online Backup 11/03/2010
First Annual Green It Economic Summit to Set New Directions for Green IT in Private and Public Secto 11/03/2010
BrainPulse Technologies Continues to Receive Popularity and Acclamation 10/03/2010
Two Models of Microsoft Phones Exposed and Posed a Challenge to iPhone and Google Nexus One 10/03/2010
Smart Mobile Phones Increased 23.8% Last Year and Touch Screen Phone Proportion Rose 08/03/2010
Company to Provide Free Salesperson for Every United States Based Business 07/03/2010
Company to Provide Free Salesperson for Every United States Based Business 07/03/2010
Radiology PACS Workstation Provider Intelerad Partners with Largest US Private Radiology Group 07/03/2010
Web Hosting Company Webhostingpad.com Introduces Unbeatable $1.99 Monthly Hosting Plans 06/03/2010
Jan 27,2010 - SEO Directory Submission Software Standard Edition Launched 05/03/2010
Jan 27,2010 - SEO Directory Submission Software Standard Edition Launched 05/03/2010
GTSS Inc. Redefines its Customized Staffing Solutions for Technology Companies in North America 05/03/2010
Jan 27,2010 - SEO Press release(PR) Submission Software Standard Edition Launched 05/03/2010
Jan 27,2010 - SEO Press release(PR) Submission Software Standard Edition Launched 05/03/2010
Poland Braced for Invasion – Izzicarhire Set to Overrun the Country With Great Value Car Hire Deals 04/03/2010
Allied Pickfords China Opens New Branch in Chengdu China 04/03/2010
Jan 27,2010 - SEO PAD Submission Software Standard Edition Launched 04/03/2010
Jan 27,2010 - SEO PAD Submission Software Standard Edition Launched 04/03/2010
Jan 27,2010 - SEO PAD Submission Software Standard Edition Launched 04/03/2010
Fashion LED Bracelet Watches Lead You a Stylish Life 04/03/2010
Thecus N4100PRO 4TB Entry Level NAS server for $699.00 03/03/2010
Izzicarhire Declares War on Over Charging 03/03/2010
PriceGo Gives Special Offerings on PS3 Controllers, and Cheap MP3 Players 03/03/2010
ePathChina Help You Have Spy Equipments Jackie Chan Using in The Spy Next Door 02/03/2010
Microland in global partnership with ChangeBASE 02/03/2010
Izzicarhire Quick to Spot Tourism Trends – Visitors to Ireland Next in Line to Benefit 28/02/2010
Sprint Electric Select Direct-Industrial.com as Distributor for DC Drives 28/02/2010
NovaBACKUP xSP Enables ISO/IEC 20000 Certification 27/02/2010
Elmasonic E Ultrasonic Cleaner with Distinctive Digital Sweep Capability 27/02/2010
Earn Real Cash From Logodesignnewzealand 26/02/2010
Firm Reveals the Real Costs in Search Engine Optimization, Fast SEO Ranking 25/02/2010
SMA & Synergent Celebrate the Automation of 10 Million Jobs 24/02/2010
Remote Hybrid Power Station Containerized 24/02/2010
Elmasonic S Ultrasonic Cleaners are ideal for parts cleaning, solvent degassing for content testing 23/02/2010
Kern Digital Scales and Industrial Lab Balances Available from Tovatech 23/02/2010
Laboratory Refrigerator is Final Link in Critical Cold Chain Storage of Vaccines 23/02/2010
Elmasonic S150 Ultrasonic Cleaner for Labs, Medical and Surgical Instruments 23/02/2010
HPLC Flowmeter from Tovatech provides Fast, Accurate Calibration and IQ OQ PQ Validation for FDA 23/02/2010
High-Capacity Industrial Ultrasonic Cleaning Equipment from Tovatech 23/02/2010
Elmasonic E Ultrasonic Cleaner with Distinctive Digital Sweep Capability 22/02/2010
MetricStream to Showcase its GRC Solutions at the Utilities & Energy Compliance & Ethics Conference 20/02/2010
MetricStream to Host a Complimentary Webinar on Operational Risk Management and Basel II Compliance 20/02/2010
Best Blog for Tech Gadgets Blueunplugged has the Latest Updates on Gadgets 18/02/2010
Synchronica Introduces Next-Generation Messaging, Adding Mobile Social Networking 18/02/2010
Print Manager Plus 2010 Now Supports Microsoft XPS 18/02/2010
Lakeville Area Website Company Owner Advises Businesses on Social Media 17/02/2010
Free Fax To Email Service Launched In The UK 15/02/2010
Online Login Helper helps surfers avoid phishing attacks 14/02/2010
Research Report Ranks 16 Global Content Management Systems 14/02/2010
Free Software for Basic PDF Needs - Multilanguage Versions with up to Nine Languages Available 14/02/2010
MJPC Member Receives Prestigious Anti-Genocide Fellowship 07/02/2010
Teach you how to let moment become eternal 04/02/2010
Wyandotte computer repair 48192, Downriver Computer Repair keeps businesses running smoothly 03/02/2010
Managing Business through Latest Mobile POS Terminals 02/02/2010
Social media campaign will expose internal sex trafficking of young teenage girls 02/02/2010
Get your free iPad - at Gadgets4Nowt free gift website 02/02/2010
The Latest Super Gorgeous MP5 Player – A Superior Music Game Machine 01/02/2010
Wipro Infotech is the Indian IT Company of the Year 2009 01/02/2010
Jobs Claimed Google Attempting Killing iPhone is Wishful Thinking 01/02/2010
ePathChina Launches Brand-New Upgraded Copy iPod NANO MP4 Player 30/01/2010
MetricStream to Host a Complimentary Webinar on Linking Internal Audit 29/01/2010
Easyspace Improves Website for non UK Visitors Buying Domain Names 28/01/2010
Esker on Demand Automates GFE Supplier Invoicing 25/01/2010
Motorola Permits Chinese Users to Change Android Phones Search Engine 25/01/2010
Contact Information
Mark Won
Labthink Instruments Co., Ltd.
+86 (0)531 85864214-8058
Website
http://www.labthink.cn/


Email
secretary@labthink.cn


Press release files
There are no files associated with this release.
Please do not contact SANEPR with questions, queries or comments regarding any information contained within press releases, distributed by or appearing on SANEPR.com Please contact the company listed in the press release as SANEPR will be not be able to assist with any such inquiry as stated in the terms of service.